SPC line: Small Punch Test machines for 650 °C to 1200 °C

Select the temperature class

Small Punch Test machine · EN 10371 · highest temperature class

SPC 1200

Creep, tensile and fracture tests up to 1200 °C, on a disc 8 mm in diameter.

The SPC 1200 does Small Punch tests from room temperature to 1200 °C, in argon, with a constant force to 5000 N. The test runs automatically, from the start command to the cool-down.

SPC 1200 Small Punch Test machine
1200 °C
maximum test temperature
5000 N
constant force, direct loading
±0.2 °C
temperature stability on the test piece
Ø8 · Ø3 mm
standard and miniature test pieces

Technical data

One platform, four temperature classes

MachineTest temperatureConstant force, direct loadingLever loading, option
SPC 650room temperature to 650 °Cto 1000 Nto 500 N
SPC 800room temperature to 800 °Cto 1000 Nto 500 N
SPC 1000room temperature to 1000 °Cto 4000 Nto 500 N
SPC 1200room temperature to 1200 °Cto 5000 Nto 500 N

All four machines do creep tests and also tensile and fracture tests, in argon, on Ø8 mm and Ø3 mm test pieces.

Select the temperature class

SPC 1200

Test types
creep · tensile and fracture
Test temperature
room temperature to 1200 °C
Temperature stability on the test piece
±0.2 °C
Temperature measuring chain
±1 °C or better
Constant force, direct loading
to 5000 N
Lever loading, option
to 500 N
Force accuracy
±1 % of the test force, from 200 N · EN ISO 7500-1, class 1
Deflection measurement
±0.001 mm · EN ISO 9513, class 1
Crosshead rate, tensile and fracture test
0.001 to 1 mm/min
Test pieces
Ø8 × 0.5 mm and Ø3 × 0.25 mm
Atmosphere
argon
Test duration
no limit
Data export
XLS · CEN XML (EN 10371, Annex I)
Test standard
EN 10371:2021

Loading

Direct loading to 5000 N is the standard, and we recommend it. Lever loading to 500 N is available as an option.

Cartridges

One cartridge for the Ø8 × 0.5 mm test piece and one for the Ø3 × 0.25 mm test piece. You can add the second cartridge later.

Developed with research partners

TREND programme of the Technology Agency of the Czech Republic, project FW06010572, with the Institute of Physics of Materials of the Czech Academy of Sciences.

Customised for your test

Tell us your material, temperature and force. We adapt the configuration of the machine.

The leaflet has the full technical data. Tell us who you are and we send it to you.

Ask for the leaflet

Why the SPC line

What the machine does for you

01

The test runs automatically

Insert the test piece, set the temperature and the force, and push Start. The machine does the purge, the heating, the loading, the recording and the cool-down.

02

The temperature is measured on the test piece

The control uses the temperature of the test piece, not the temperature of the furnace. The machine measures to ±1 °C or better and keeps the temperature stable to ±0.2 °C. EN 10371 permits ±3 °C at 800 °C.

03

The force comes without shock

The machine applies the force automatically and smoothly. Then it keeps the force constant for the full test.

04

Argon protects the test piece

The test is in argon, in a sealed chamber. This protects the thin test piece from oxidation.

05

A power failure does not stop the test

A UPS keeps the machine in operation during a short power failure. A long creep test continues.

06

Data that you can use

The machine records temperature, force and deflection for the full test. You export the data as XLS or in the CEN XML format of EN 10371, Annex I.

Repeatability

Three test pieces, one curve

Small punch creep tests at 600 °C: three tests for each force, all to rupture.

Small punch creep curves at 600 °C: three tests at 450 N and three at 520 N
≈ 20 h

450 N · 3 tests. Range less than 0.5 h, approximately ±1.3 %.

≈ 8 h

520 N · 3 tests. Range less than 0.5 h, approximately ±3 %.

±0.1 °C

On the test piece for 62 h at 600 °C. Published result for the SPC 1300 DLS, the highest class of the same platform (ECCC 2026).

Delivery

Complete, with training

In the delivery

  • Machine, control unit and computer with the control software.
  • Cartridges for your test pieces.
  • Reference test pieces. You can check the machine at any time.
  • Two days of operator training in your laboratory, on your machine and with your test pieces.
  • Accessories and spare parts.

Your laboratory

Machine
780 × 780 × 2040 mm · 130 kg
With the test table
1800 × 1000 × 2040 mm · 185 kg
Power supply
230 V AC, 16 A
Gas
argon supply
Room
air-conditioned, 23 ± 2 °C

Do you want to see results first?

Send us your material. We do Small Punch tests as a service on our machines, and you get the test report.

Ask for a test

Interested in an SPC machine?

Tell us which materials and temperatures you need to cover — we will propose a configuration and a delivery time.

Request more information

Data protection notice

Information on data processing can be found in our privacy statement.

Mandatory