Small Punch Test · +23 to −193 °C · EN 10371:2021
TTS 190
Cryogenic Small Punch Test system, supplied complete with its evaluation software.
Fill the Dewar with liquid nitrogen, insert the test piece and set the temperature — cooling, the test and the evaluation to EN 10371 run automatically, from room temperature down to −193 °C.
−193 °C
lowest test temperature (LN₂)
±1 °C
measured on the test piece surface
±1 µm
deflection measurement
10 kN
load range, accuracy class 0.05
Where this machine sits in the temperature range
What is measured and how it is evaluated
Two quantities are measured. SPEVAL calculates everything else from them automatically, with the compliance of the testing machine taken into account. The references are to EN 10371:2021.
| Quantity | Symbol | How it is obtained | EN 10371 |
|---|---|---|---|
| Force-deflection curve | F(u) | Measured. Deflection u on the underside of the test piece, ±1 µm; punch displacement v is recorded as well. | Clause 7 |
| Test temperature | T | Measured on the test piece surface, not in the chamber. Held to ±1 °C, resolution 0.1 °C. | Annex B |
| 0.2 % proof strength | Rp0.2 | From the elastic-plastic transition force Fe, by the bilinear method (or manually with two tangents). | Annex D |
| Tensile strength | Rm | Equation C.1 or C.2, whichever applies to the material. | Annex C |
| Small punch energy | ESP | Area under F(u) up to um, the deflection at maximum force Fm — or up to the first significant pop-in (a force drop of at least 0.1 Fm). | Annex E |
| Small punch transition temperature | TSP | Least-squares tanh fit of ESP/Fm against T; TSP is the midpoint between upper and lower shelf energy. | Annex E |
| DBTT estimate | — | Converted from TSP by correlation, TSP = α · TCVN, both in kelvin. | Annex E |
A transition curve from one material batch
Each point is one test piece at one temperature. Normalising ESP by Fm keeps the upper shelf flat although strength rises as the steel gets colder, so a single tanh function fits the whole range.
SPEVAL builds the curve from the measured series and determines TSP from it — without fitting the curve by hand.
Technical parameters
- Temperature range
- +23 to −193 °C (higher temperatures as a custom adjustment)
- Temperature control
- ±1 °C, resolution 0.1 °C, on the test piece surface
- Load range
- up to 10 kN, accuracy class 0.05
- Deflection measurement
- ±1 µm
- Test piece
- Ø8 × 0.5 mm (optionally 10 × 10 or 8 × 8 × 0.5 mm)
- Punch
- Ø2.5 mm (optionally Ø2 mm)
- Lower die
- Ø4 mm receiving hole, R0.2 (optionally 0.2 × 45° chamfer)
- Liquid nitrogen
- 32 l Dewar vessel, automatic dosing
- Evaluated quantities
- Rp0.2 · Rm · ESP · TSP · DBTT estimate
- Standard
- EN 10371:2021
System composition
Supplied complete: frame, cryogenics, control, PC and evaluation software — nothing to buy on top. You fill the Dewar, insert the test piece and set the temperature; the system does the rest. Verified with reference test pieces supplied with it.
| Component | Function | Size · mass |
|---|---|---|
| Universal testing machine | 10 kN frame, accuracy class 0.05 | 1600 × 600 × 720 mm · 180 kg |
| Test mandrel | Clamps the test piece between upper and lower die; carries the deflection and temperature sensors | 450 × Ø120 mm |
| Cryogenic chamber | Insulated, LN₂ inlet, internal sensor, heater for trimming | 250 × 190 × 250 mm · 15 kg |
| Dosing probe | Pressurises the Dewar and doses LN₂ automatically; relief and solenoid valves | 840 × 200 × 150 mm |
| Dewar vessel | LN₂ supply for a test day | 32 l · Ø500 × 725 mm · 20 kg |
| Control units, PC and SPEVAL | Cooling, heating, nitrogen dosing and test control; evaluation to EN 10371 | 7 + 7 kg · PC 11 kg |
Footprint with Dewar 2200 × 700 × 2320 mm · power 1 NPE 220–230 V, 10 A · chamber for higher temperatures as a custom adjustment, 15 kg.
Included
SPEVAL
Evaluation software to EN 10371, installed on the supplied PC. The same calculation, the same way, every time.
- Machine compliance correction Results are calculated with the compliance of the testing machine taken into account, so they compare between laboratories.
- Rp0.2 automatically or manually Automatic bilinear method or manual two tangents, from deflection or from punch displacement.
- Rm to the standard Equation C.1 or C.2, whichever applies to the material.
- Energy with data export ESP calculated automatically and exported for the test report.
- TSP and DBTT Determined from the measured series, without fitting the curve by hand.
- Reads other data too Any test file, including tests run elsewhere or earlier.
Send us the material and the question
Steel grade and condition, the temperature range of interest and where the test pieces will come from. We will come back with a configuration — and, for a service-exposed component, with a sampling proposal too.
Product leaflet
A leaflet with more information about the product. This is only available upon request.