Small Punch Test · +23 to −193 °C · EN 10371:2021

TTS 190

Cryogenic Small Punch Test system, supplied complete with its evaluation software.

Fill the Dewar with liquid nitrogen, insert the test piece and set the temperature — cooling, the test and the evaluation to EN 10371 run automatically, from room temperature down to −193 °C.

TTS 190

−193 °C

lowest test temperature (LN₂)

±1 °C

measured on the test piece surface

±1 µm

deflection measurement

10 kN

load range, accuracy class 0.05

What is measured and how it is evaluated

Two quantities are measured. SPEVAL calculates everything else from them automatically, with the compliance of the testing machine taken into account. The references are to EN 10371:2021.

Quantity Symbol How it is obtained EN 10371
Force-deflection curve F(u) Measured. Deflection u on the underside of the test piece, ±1 µm; punch displacement v is recorded as well. Clause 7
Test temperature T Measured on the test piece surface, not in the chamber. Held to ±1 °C, resolution 0.1 °C. Annex B
0.2 % proof strength Rp0.2 From the elastic-plastic transition force Fe, by the bilinear method (or manually with two tangents). Annex D
Tensile strength Rm Equation C.1 or C.2, whichever applies to the material. Annex C
Small punch energy ESP Area under F(u) up to um, the deflection at maximum force Fm — or up to the first significant pop-in (a force drop of at least 0.1 Fm). Annex E
Small punch transition temperature TSP Least-squares tanh fit of ESP/Fm against T; TSP is the midpoint between upper and lower shelf energy. Annex E
DBTT estimate — Converted from TSP by correlation, TSP = α · TCVN, both in kelvin. Annex E

A transition curve from one material batch

Each point is one test piece at one temperature. Normalising ESP by Fm keeps the upper shelf flat although strength rises as the steel gets colder, so a single tanh function fits the whole range.

SPEVAL builds the curve from the measured series and determines TSP from it — without fitting the curve by hand.

TSP = −140 °C upper shelf lower shelf −200−150−100−500 test temperature T (°C) ESP / Fm
Illustrative data.

Technical parameters

Temperature range
+23 to −193 °C (higher temperatures as a custom adjustment)
Temperature control
±1 °C, resolution 0.1 °C, on the test piece surface
Load range
up to 10 kN, accuracy class 0.05
Deflection measurement
±1 µm
Test piece
Ø8 × 0.5 mm (optionally 10 × 10 or 8 × 8 × 0.5 mm)
Punch
Ø2.5 mm (optionally Ø2 mm)
Lower die
Ø4 mm receiving hole, R0.2 (optionally 0.2 × 45° chamfer)
Liquid nitrogen
32 l Dewar vessel, automatic dosing
Evaluated quantities
Rp0.2 · Rm · ESP · TSP · DBTT estimate
Standard
EN 10371:2021

System composition

Supplied complete: frame, cryogenics, control, PC and evaluation software — nothing to buy on top. You fill the Dewar, insert the test piece and set the temperature; the system does the rest. Verified with reference test pieces supplied with it.

Component Function Size · mass
Universal testing machine 10 kN frame, accuracy class 0.05 1600 × 600 × 720 mm · 180 kg
Test mandrel Clamps the test piece between upper and lower die; carries the deflection and temperature sensors 450 × Ø120 mm
Cryogenic chamber Insulated, LN₂ inlet, internal sensor, heater for trimming 250 × 190 × 250 mm · 15 kg
Dosing probe Pressurises the Dewar and doses LN₂ automatically; relief and solenoid valves 840 × 200 × 150 mm
Dewar vessel LN₂ supply for a test day 32 l · Ø500 × 725 mm · 20 kg
Control units, PC and SPEVAL Cooling, heating, nitrogen dosing and test control; evaluation to EN 10371 7 + 7 kg · PC 11 kg

Footprint with Dewar 2200 × 700 × 2320 mm · power 1 NPE 220–230 V, 10 A · chamber for higher temperatures as a custom adjustment, 15 kg.

Included

SPEVAL

Evaluation software to EN 10371, installed on the supplied PC. The same calculation, the same way, every time.

  • Machine compliance correction Results are calculated with the compliance of the testing machine taken into account, so they compare between laboratories.
  • Rp0.2 automatically or manually Automatic bilinear method or manual two tangents, from deflection or from punch displacement.
  • Rm to the standard Equation C.1 or C.2, whichever applies to the material.
  • Energy with data export ESP calculated automatically and exported for the test report.
  • TSP and DBTT Determined from the measured series, without fitting the curve by hand.
  • Reads other data too Any test file, including tests run elsewhere or earlier.

Send us the material and the question

Steel grade and condition, the temperature range of interest and where the test pieces will come from. We will come back with a configuration — and, for a service-exposed component, with a sampling proposal too.

Product leaflet — TTS 190

Product leaflet

A leaflet with more information about the product. This is only available upon request.